A Self-Calibrated Transfer Standard For Microwave Calorimetry

2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018)(2018)

引用 1|浏览6
暂无评分
摘要
We developed a new calibration technique for measuring the correction factor of a calorimeter with a vector network analyzer. Based on a wave-parameter formulation, we developed analytic formulas for the correction-factor (g) and effective-efficiency (eta). This allows us to calibrate both the calorimeter and the thermistor mount in a single step. This greatly reduces the number of physical parameters involved in calibration processes while tracking correlated uncertainty.
更多
查看译文
关键词
Correction factor, microcalorimeter, microwave power, transfer standard, vector wave parameter