An X-ray gas monitor for free-electron lasers.

JOURNAL OF SYNCHROTRON RADIATION(2019)

引用 32|浏览73
暂无评分
摘要
A novel X-ray gas monitor (XGM) has been developed which allows the measurement of absolute photon pulse energy and photon beam position at all existing and upcoming free-electron lasers (FELs) over a broad spectral range covering vacuum ultraviolet (VUV), extreme ultraviolet (EUV) and soft and hard X-rays. The XGM covers a wide dynamic range from spontaneous undulator radiation to FEL radiation and provides a temporal resolution of better than 200ns. The XGM consists of two X-ray gas-monitor detectors (XGMDs) and two huge-aperture open electron multipliers (HAMPs). The HAMP enhances the detection efficiency of the XGM for low-intensity radiation down to 10(5) photons per pulse and for FEL radiation in the hard X-ray spectral range, while the XGMD operates in higher-intensity regimes. The relative standard uncertainty for measurements of the absolute photon pulse energy is well below 10%, and down to 1% for measurements of relative pulse-to-pulse intensity on pulses with more than 10(10) photons per pulse. The accuracy of beam-position monitoring in the vertical and horizontal directions is of the order of 10 mu m.
更多
查看译文
关键词
free-electron lasers,vacuum ultraviolet,soft X-rays,hard X-rays,photon diagnostics