Evaluation of an in-house self-held respiration monitoring device for deep inspiration breath hold techniques for radiotherapy of patients with cancer of the left breast

Yuen Yan Chan, Ki Man Ku, Yin Ping Ng,Siu Ki Ben Yu,Chi Wah Tony Kong, Wing Lun Alan Mui, Chun Man Cornel Wong, Sze Wan Kong, Sze Ming Wong,George Chiu

JOURNAL OF RADIOTHERAPY IN PRACTICE(2020)

引用 0|浏览0
暂无评分
摘要
Background: An in-house self-held respiration monitoring device (SHRMD) was developed for providing deep inspiration breath hold (DIBH) radiotherapy. The use of SHRMD is evaluated in terms of reproducibility, stability and heart dose reduction. Methods and materials: Sixteen patients receiving radiotherapy of left breast cancer were planned for treatment with both a free breathing (FB) scan and a DIBH scan. Both FB and DIBH plans were generated for comparison of the heart, left anterior descending (LAD) artery and lung dose. All patients received their treatments with DIBH using SHRMD. Megavoltage cine images were acquired during treatments for evaluating the reproducibility and stability of treatment position using SHRMD. Results: Compared with FB plans, the maximum dose to the heart by DIBH technique with SHRMD was reduced by 29 center dot 9 +/- 15 center dot 6%; and the maximum dose of the LAD artery was reduced by 41 center dot 6 +/- 18 center dot 3%. The inter-fractional overall mean error was 0 center dot 01 cm and the intra-fractional overall mean error was 0 center dot 04 cm. Conclusion: This study demonstrated the potential benefits of using the SHRMD for DIBH to reduce the heart and LAD dose. The patients were able to perform stable and reproducible DIBHs.
更多
查看译文
关键词
breath-hold,heart dose,reproducibility,SHRMD,stability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要