A Novel Hybrid 3d Endoscope Zooming And Repositioning System: Design And Feasibility Study

INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY(2020)

引用 0|浏览25
暂无评分
摘要
Background Manipulation of the endoscope during minimally invasive surgery is a major source of inconvenience and discomfort. This report elucidates the architecture of a novel one-hand controlled endoscope positioning device and presents a practicability evaluation. Methods and materials Setup time and total surgery time, number and duration of the manipulations, side effects of three-dimensional (3D) imaging, and ergonomic complaints were assessed by three surgeons during cadaveric and in vivo porcine trials. Results Setup was accomplished in an average (SD) of 230 (120) seconds. The manipulation time was 3.87 (1.77) seconds for angular movements and 0.83 (0.24) seconds for zooming, with an average (SD) of 30.5 (16.3) manipulations per procedure. No side effects of 3D imaging or ergonomic complaints were reported. Conclusions The integration of an active zoom into a passive endoscope holder delivers a convenient synergy between a human and a machine-controlled holding device. It is shown to be safe, simple, and intuitive to use and allows unrestrained autonomic control of the endoscope by the surgeon.
更多
查看译文
关键词
endoscope positioner, minimally invasive surgery, solo surgery, surgical assistant