Practical implementation of robust MR-thermometry during clinical MR-guided microwave ablations in the liver at 1.5 T.

Physica Medica(2019)

引用 19|浏览22
暂无评分
摘要
•Practical measures to achieve robust real-time MR thermometry during clinical hepatic microwave ablations are described.•Quality Assurance protocol was developed ensuring consistent high quality imaging during MR-guided microwave procedures.•Imaging protocol was modified to account for respiratory movement and reduce respiratory motion artifact.•These measures consistently reduced image noise and artifacts improving safety and effectiveness of microwave treatments.
更多
查看译文
关键词
MR-guided microwave ablations,MR-thermometry,Quality assurance,Artifact reduction,Interventional MRI