Impact of Corneal Parameters on Intraocular Pressure Measurements in Different Tonometry Methods

International journal of ophthalmology(2019)

引用 14|浏览14
暂无评分
摘要
AIM: To evaluate the impact of central corneal thickness (CCT) and corneal curvature on intraocular pressure (IOP) measurements performed by three different tonometers. METHODS: IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), and ICare rebound tonometer (RT). CCT and corneal curvature were assessed. RESULTS: In healthy eyes, DCT presents significantly higher values of IOP than GAT (17.34 +/- 3.69 and 15.27 +/- 4.06 mm Hg, P<0.0001). RT measurements are significantly lower than GAT (13.56 +/- 4.33 mm Hg, P<0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 mu m and 0.99 mm Hg higher results in eyes with CCT >= 600 mu m. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 mu m were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 mu m was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values (R=0.369, P=0.005). CONCLUSION: The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT; thus, CCT should be taken into consideration for both diagnostics and monitoring.
更多
查看译文
关键词
intraocular pressure,Goldmann applanation tonometer,Pascal dynamic contour tonometer,ICare rebound tonometer,central corneal thickness,corneal curvature,healthy individuals
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要