MOSFET dosimeter characterization in MR-guided radiation therapy (MRgRT) Linac.

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS(2020)

引用 4|浏览37
暂无评分
摘要
Purpose With the increasing use of MR-guided radiation therapy (MRgRT), it becomes important to understand and explore accuracy of medical dosimeters in the presence of magnetic field. The purpose of this work is to characterize metal-oxide-semiconductor field-effect transistors (MOSFETs) in MRgRT systems at 0.345 T magnetic field strength. Methods A MOSFET dosimetry system, developed by Best Medical Canada for in-vivo patient dosimetry, was used to study various commissioning tests performed on a MRgRT system, MRIdian (R) Linac. We characterized the MOSFET dosimeter with different cable lengths by determining its calibration factor, monitor unit linearity, angular dependence, field size dependence, percentage depth dose (PDD) variation, output factor change, and intensity modulated radiation therapy quality assurance (IMRT QA) verification for several plans. MOSFET results were analyzed and compared with commissioning data and Monte Carlo calculations. Results MOSFET measurements were not found to be affected by the presence of 0.345 T magnetic field. Calibration factors were similar for different cable length dosimeters either placed at the parallel or perpendicular direction to the magnetic field, with variations of less than 2%. The detector showed good linearity (R-2 = 0.999) for 100-600 MUs range. Output factor measurements were consistent with ionization chamber data within 2.2%. MOSFET PDD measurements were found to be within 1% for 1-15 cm depth range in comparison to ionization chamber. MOSFET normalized angular response matched thermoluminescent detector (TLD) response within 5.5%. The IMRT QA verification data for the MRgRT linac showed that the percentage difference between ionization chamber and MOSFET was 0.91%, 2.05%, and 2.63%, respectively for liver, spine, and mediastinum. Conclusion MOSFET dosimeters are not affected by the 0.345 T magnetic field in MRgRT system. They showed physics parameters and performance comparable to TLD and ionization chamber; thus, they constitute an alternative to TLD for real-time in-vivo dosimetry in MRgRT procedures.
更多
查看译文
关键词
beam QA,dosimetry,IMRT,low magnetic field,MOSFET,MRgRT
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要