Optimal ECG Lead System for Exercise Assessment of Ischemic Heart Disease
JOURNAL OF CARDIOVASCULAR TRANSLATIONAL RESEARCH(2019)
摘要
The diagnostic value of an ECG exercise test in diagnosis of ischemic heart disease (IHD) is limited. We investigated whether it is possible to develop a method for diagnosis of IHD which uses a low number of optimal ECG leads and has a higher diagnostic efficiency than conventional exercise ECG. This study was carried out on 43 patients. The 67-lead high-resolution ECG was recorded at rest and during exercise. The diagnostic value of ST segment depression (ΔST60) and T-wave morphology change (δT) determined in optimized ECG lead configurations was higher than for the standard 12-lead ECG. The best results were obtained for δT determined from 6 ECG electrode locations where sensitivity and specificity were 70% and 69% whereas for the standard exercise ECG were 63% and 62%, respectively. The small number of ECG leads used allows for easy hardware implementation of the methods for use in clinical settings.
更多查看译文
关键词
Multi-lead high-resolution ECG,Ischemic heart disease,Exercise test,Optimal ECG leads system