On Low-Energy Tail Distortions in the Detector Response Function of X-Ray Microcalorimeter Spectrometers

Journal of Low Temperature Physics(2019)

引用 9|浏览93
暂无评分
摘要
We use narrow spectral lines from the X-ray spectra of various highly charged ions to measure low-energy tail-like deviations from a Gaussian response function in a microcalorimeter X-ray spectrometer with Au absorbers at energies from 650 to 3320 eV. We review the literature on low-energy tails in other microcalorimeter X-ray spectrometers and present a model that explains all the reviewed tail fraction measurements. In this model, a low-energy tail arises from the combination of electron escape and energy trapping associated with Bi X-ray absorbers.
更多
查看译文
关键词
Transition-edge sensor,Low-energy tail,Point spread function,Microcalorimeter,JMONSEL