Pivot Calibration Concept For Sensor Attached Mobile C-Arms
MEDICAL IMAGING 2020: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING(2021)
摘要
Medical augmented reality has been actively studied for decades and many methods have been proposed to revolutionize clinical procedures. One example is the camera augmented mobile C-arm (CAMC), which provides a real-time video augmentation onto medical images by rigidly mounting and calibrating a camera to the imaging device. Since then, several CAMC variations have been suggested by calibrating 2D/3D cameras, trackers, and more recently a Microsoft HoloLens to the C-arm. Different calibration methods have been applied to establish the correspondence between the rigidly attached sensor and the imaging device. A crucial step for these methods is the acquisition of X-Ray images or 3D reconstruction volumes; therefore, requiring the emission of ionizing radiation. In this work, we analyze the mechanical motion of the device and propose an alternatative method to calibrate sensors to the C-arm without emitting any radiation. Given a sensor is rigidly attached to the device, we introduce an extended pivot calibration concept to compute the fixed translation from the sensor to the C-arm rotation center. The fixed relationship between the sensor and rotation center can be formulated as a pivot calibration problem with the pivot point moving on a locus. Our method exploits the rigid C-arm motion describing a Torus surface to solve this calibration problem. We explain the geometry of the C-arm motion and its relation to the attached sensor, propose a calibration algorithm and show its robustness against noise, as well as trajectory and observed pose density by computer simulations. We discuss this geometric-based formulation and its potential extensions to different C-arm applications.
更多查看译文
关键词
Calibration, Pivot Calibration, Torus, Augmented Reality, Mobile C-arm, CAMC
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要