Study of Position Sensitive Silicon Detector (PSD) for SiW-ECAL at ILC

Uesugi Y.,Mori R., Yamashiro H., Suehara T.,Yoshioka T., Kawagoe K.

JOURNAL OF INSTRUMENTATION(2020)

引用 0|浏览14
暂无评分
摘要
For the past few years we have been developing position sensitive silicon detectors (PSDs) which have an electrode at each of four corners so that incident position of a charged particle can be obtained with signal from the electrodes. It is expected that the position resolution of the silicon-tungsten electromagnetic calorimeter (SiW-ECAL) of the International Large Detector (ILD) will be improved by introducing PSDs to detection layers. In the previous production we found that the charge separation is not optimal due to the readout impedance. To solve the issue, we produced new PSDs with higher surface resistance with an additional resistive layer on the surface. We also implemented several techniques to decrease position distortion and increase signal-to-noise ratio which are essential for optimal position resolution. We present first measurements of the performance of one new PSD using a strontium 90 source, and using the Skiroc2-CMS ASIC.