Numerical method for axial motion artifact correction in retinal spectral-domain optical coherence tomography

FRONTIERS OF OPTOELECTRONICS(2019)

引用 6|浏览52
暂无评分
摘要
A numerical method that compensates image distortions caused by random fluctuations of the distance to an object in spectral-domain optical coherence tomography (SD OCT) has been proposed and verified experimentally. The proposed method is based on the analysis of the phase shifts between adjacent scans that are caused by micrometer-scale displacements and the subsequent compensation for the displacements through phase-frequency correction in the spectral space. The efficiency of the method is demonstrated in model experiments with harmonic and random movements of a scattering object as well as during in vivo imaging of the retina of the human eye.
更多
查看译文
关键词
optical coherence tomography (OCT), motion artifact correction, retinal imaging, numerical method
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper