谷歌浏览器插件
订阅小程序
在清言上使用

New advances in mechanomyography sensor technology and signal processing: Validity and intrarater reliability of recordings from muscle

JOURNAL OF REHABILITATION AND ASSISTIVE TECHNOLOGIES ENGINEERING(2020)

引用 7|浏览20
暂无评分
摘要
Introduction The Mechanical Muscle Activity with Real-time Kinematics project aims to develop a device incorporating wearable sensors for arm rehabilitation following stroke. These will record kinematic activity using inertial measurement units and mechanical muscle activity. The gold standard for measuring muscle activity is electromyography; however, mechanomyography offers an appropriate alterative for our home-based rehabilitation device. We have patent filed a new laboratory-tested device that combines an inertial measurement unit with mechanomyography. We report on the validity and reliability of the mechanomyography against electromyography sensors. Methods In 18 healthy adults (27-82 years), mechanomyography and electromyography recordings were taken from the forearm flexor and extensor muscles during voluntary contractions. Isometric contractions were performed at different percentages of maximal force to examine the validity of mechanomyography. Root-mean-square of mechanomyography and electromyography was measured during 1 s epocs of isometric flexion and extension. Dynamic contractions were recorded during a tracking task on two days, one week apart, to examine reliability of muscle onset timing. Results Reliability of mechanomyography onset was high (intraclass correlation coefficient = 0.78) and was comparable with electromyography (intraclass correlation coefficient = 0.79). The correlation between force and mechanomyography was high (R-2 = 0.94). Conclusion The mechanomyography device records valid and reliable signals of mechanical muscle activity on different days.
更多
查看译文
关键词
Rehabilitation,sensors,sensor applications,rehabilitation devices,upper-limb,electromyography,sensor design,mechanomyography
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要