Comparison of radial and meander-like breast ultrasound with respect to diagnostic accuracy and examination time

Archives of Gynecology and Obstetrics(2020)

引用 2|浏览9
暂无评分
摘要
Purpose To prospectively compare the diagnostic accuracy of radial breast ultrasound (r-US) to that of conventional meander-like breast ultrasound (m-US), patients of a consecutive, unselected, mixed collective were examined by both scanning methods. Methods Out of 1948 dual examinations, 150 revealed suspicious lesions resulting in 168 biopsies taken from 148 patients. Histology confirmed breast cancers in 36 cases. Sensitivity, specificity, accuracy, PPV, and NPV were calculated for r-US and m-US. The examination times were recorded. Results For m-US and r-US, sensitivity (both 88.9%), specificity (86.4% versus 89.4%), accuracy (86.9% versus 89.3%), PPV (64.0% versus 69.6%), NPV (both 98.3%), false-negative rate (both 5.6%), and rate of cancer missed by one method (both 5.6%) were similar. The mean examination time for r-US (14.8 min) was significantly ( p < 0.01) shorter than for m-US (22.6 min). Conclusion Because the diagnostic accuracy of r-US and m-US are comparable, r-US can be considered an alternative to m-US in routine breast US with the added benefit of a significantly shorter examination time.
更多
查看译文
关键词
Diagnostic accuracy, Ductosonography, Examination time, False negative, Meander-like breast ultrasound, Radial breast ultrasound