Non-contact Direct Selective Laser Trabeculoplasty: Light Propagation Analysis.

Biomedical optics express(2020)

引用 4|浏览11
暂无评分
摘要
Selective laser trabeculoplasty (SLT), used to treat glaucoma and ocular hypertension, requires the use of a gonioscope placed on the cornea to visualize and irradiate the trabecular meshwork (TM). Alternatively, non-contact direct SLT (DSLT) irradiates the TM through the overlying tissues. Here we analyze this innovative procedure using analytical modeling and Monte Carlo simulations to quantify the laser energy reaching the TM through the overlying tissues. Compared with energy launched from the laser, DSLT energy transmission to the TM is 2.8 times less than SLT, which verifies the efficacy of non-contact DSLT given the lowest reported effective SLT energies.