Multi-Laboratory Validation Of The Xmap-Food Allergen Detection Assay: A Multiplex, Antibody-Based Assay For The Simultaneous Detection Of Food Allergens

PLOS ONE(2020)

引用 12|浏览21
暂无评分
摘要
The increasing prevalence of individuals with multiple food allergies and the need to distinguish between foods containing homologous, cross-reactive proteins have made the use of single-analyte antibody-based methods (e.g., ELISAs) sometimes insufficient. These issues have resulted in the need to conduct multiple analyses and sometimes employ orthogonal methods like mass spectrometry or DNA-based methods for confirmatory purposes. The xMAP Food Allergen Detection Assay (xMAP FADA) was developed to solve this problem while also providing increased throughput and a modular design suitable for adapting to changes in analytical needs. The use of built-in redundancy provides the xMAP FADA with built-in confirmatory analytical capability by including complementary antibody bead sets and secondary analytical end points (e.g., ratio analysis and multi-antibody profiling). A measure of a method's utility is its performance when employed by analysts of varying expertise in multiple laboratory environments. To gauge this aspect, a multi-laboratory validation (MLV) was conducted with 11 participants of different levels of proficiency. The MLV entailed the analysis of incurred food samples in four problematic food matrices, meat sausage, orange juice, baked muffins, and dark chocolate. Except for a couple of instances, involving two confirmatory components in the analysis of baked muffins, the allergenic foods were detected by all participants at concentrations in the analytical samples comparable to <= 10 mu g/g in the original food sample. In addition, despite high levels of inter-lab variance in the absolute intensities of the responses, the intra-laboratory reproducibility was sufficient to support analyses based on the calibration standards and direct comparison controls (DCCs) analyzed alongside the samples. In contrast, ratio analyses displayed inter-laboratory %CV (RSDR) values < 20%; presumably because the ratios are based on inherent properties of the antigenic elements. The excellent performance of the xMAP FADA when performed by analysts of varying proficiency indicates a reliability sufficient to meet analytical needs.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要