A ± 6 ms-Accuracy , 0 . 68 mm 2 , and 2 . 21 μ WQRSDetection ASIC
semanticscholar(2015)
摘要
Healthcare issues arose from population aging. Meanwhile, electrocardiogram (ECG) is a powerful measurement tool. The first step of ECG is to detect QRS complexes. A state-of-the-art QRS detection algorithm was modified and implemented to an application-specific integrated circuit (ASIC). By the dedicated architecture design, the novel ASIC is proposed with 0.68 mm2 core area and 2.21 μW power consumption. It is the smallest QRS detection ASIC based on 0.18 μm technology. In addition, the sensitivity is 95.65% and the positive prediction of the ASIC is 99.36% based on the MIT/BIH arrhythmia database certification.
更多查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要