Characterization Of Rdx And Hmx Explosive Adduct Ions Using Esi Ft-Icr Ms

JOURNAL OF MASS SPECTROMETRY(2021)

引用 1|浏览7
暂无评分
摘要
Investigation of two common explosives such as cyclonite (RDX) and cyclotetramethylenetetranitramine (HMX) using a mass spectrometer with ultrahigh resolution and accuracy has not been comprehensively performed. Here, ultrahigh mass accuracy 15-T Fourier transform-ion cyclotron resonance mass spectrometry (FT-ICR MS) spectra were utilized to comprehensively characterize the adduct ions of RDX and HMX. Two different ionization sources such as a conventional electrospray ionization (ESI) source and a chip-based static nano-ESI source were used to investigate the adduct ions of RDX and HMX. The ESI-MS analyses of two explosives in negative ion mode provide some adduct ions of RDX and HMX even without prior addition of their corresponding anions. A total of six types of adduct ion were characterized: [M + Cl](-), [M + HCOO](-), [M + NO2](-), [M + CH3COO](-), [M + NO3](-), and [M + C3H5O3](-), where M is either RDX or HMX. The ultrahigh accuracy of the 15-T FT-ICR MS was utilized to distinguish two closely spaced peaks representing the monoisotopic [M + NO2](-)and second isotopic [M + HCOO](-)ions, thereby enabling the discovery of a [M + NO2](-)adduct ion in the ESI analysis of RDX or HMX. [M + NO2](-)and [M + CH3COO](-)adduct ions were only observed when using a static nano-ESI source. It is the first report explaining the discovery of [M + NO2](-)adduct ion in the ESI-MS analyses of RDX and HMX.
更多
查看译文
关键词
ESI, explosives, FT-ICR MS, HMX, RDX