Standardization And Harmonization In Hematology: Instrument Alignment, Quality Control Materials, And Commutability Issue

INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY(2021)

引用 10|浏览15
暂无评分
摘要
Introduction In the hub and spoke laboratory network, the number of hematology analyzers (HAs) within each core center has increased, and the control of HAs alignment is becoming necessary requirement to ensure analytical quality. In this scenario, HA alignment can be assessed by analyzing the same control material used for internal quality control on multiple HAs, assuming its commutability. The aim of the study was to verify the applicability of a protocol for the alignment of HAs based on control material rather than on fresh whole-blood samples.Methods The alignment of five HAs was evaluated for red (RBC, Hb, MCV, RET), white (WBC, NE, LY, MO, EO, BA, IG), and platelet (PLT) series parameters, following a protocol by SIBioC, using human sample (HS) and quality control material (QC), after the verification of commutability, according to the IFCC protocol. Maximum bias was derived from biological variation data.Results A complete alignment between instruments was confirmed for the majority of the parameters investigated both for HS and QC material. Partial misalignments or inconcludent results were instead evident for MCV, MO, EO, BA, and IG. Interestingly, QC material was found to be not commutable for LY, MO, and BA.Conclusion The alignment of hematologic analyzers for main cell population parameters may be verified with both QC and HS, displaying consistent results and interpretation. The evaluation for some white series parameters (EO, BA, and IG) is critical, and particular attention must be paid to the values of the material used for the alignment.
更多
查看译文
关键词
alignment, analytical quality, commutability, control material, hematologic analyzers
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要