Monte-Carlo simulation of pinhole collimator of a small field of view gamma camera for small animal imaging

Zhu Jie,Ma Wenyan, Zhu Yufeng,Ma Hongguang,Wu Yuelei,Hu Huasi,Zhang Boping, Huo Yonggang, Liu Silu, Han Bin,Wang Zhaomin

NUCLEAR SCIENCE AND TECHNIQUES(2009)

引用 0|浏览2
暂无评分
摘要
Needs in scintimammography applications, especially for small animal cardiac imaging, lead to develop a small field of view, high spatial resolution gamma camera with a pinhole collimator. However the ideal pinhole collimator must keep a compromise between spatial resolution and sensitivity. In order to design a pinhole collimator with an optimized sensitivity and spatial resolution, the spatial resolution and the geometric sensitivity response as a function of the source to collimator distance has been obtained by means of Monte-Carlo simulation for a small field of view gamma camera with a pinhole collimator of various-hole diameters. The results show that the camera with pinhole of I mm, 1.5 mm and 2 min diameter has respectively spatial resolution of 1.5 mm, 2.25 min and 3 mm and geometric sensitivity of 0.016%, 0.022% and 0.036%, while the source to collimator distance is 3 cm. We chose the pinhole collimator with hole diameter size of 1.2 mm for our the gamma camera designed based on the trade-off between sensitivity and resolution.
更多
查看译文
关键词
Monte-Carlo simulation,Pinhole collimator,Gamma camera,Small field of view,High spatial resolution