A Low Noise Apd Readout Asic For Electromagnetic Calorimeter In Hiepa

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2021)

引用 5|浏览8
暂无评分
摘要
This paper presents a 4-channel avalanche photodiode readout ASIC for the electromagnetic calorimeter in the 2-7 GeV high intensity electron positron accelerator. A large detector capacitance of the avalanche photodiode will increase equivalent noise charge and extend rise time of the front-end preamplifier, resulting in attenuated energy and time resolution of the ASIC. A low-noise preamplifier and a 5th-order high-linearity shaper were proposed to decrease the noise induced by a 270 p(F) detector capacitance. The ASIC is fabricated using a standard 0.18-mu m CMOS technology, and the die size of the prototype chip is 1.1 mm x 2.4 mm. The input charge range of the ASIC is from 4.8 fC to 400 fC. The event rate of the ASIC can achieve 400k count/s. Nonlinear error of the shaper is less than 2%. The measured signal to noise ratio is better than 7.8 at a shaping time of 800 ns. The timing resolution of the ASIC is better than 300 ps at an input charge of 100 fC.
更多
查看译文
关键词
APD readout, ASIC, Low noise, Electromagnetic calorimeter