Highly sensitive determination of alendronate in human plasma and dialysate using metal-free HPLC-MS/MS

Legal Medicine(2018)

引用 0|浏览0
暂无评分
摘要
A highly sensitive method was developed for the analysis of alendronate in human plasma and dialysate using MonoSpin™ SAX® extraction and metal-free high-performance liquid chromatography (HPLC)-tandem mass spectrometry (MS/MS) following methylation with trimethylsilyldiazomethane. The chromatographic separation of the derivatives for alendronate and alendronate-d6 was achieved on an L-column2 ODS metal-free column (50 mm  ×  2 mm i.d., particle size 3 µm) with a linear gradient elution system composed of 10 mM ammonium acetate (pH 6.8) and acetonitrile at a flow rate of 0.3 ml/min. Quantification was performed by multiple reaction monitoring (MRM) with positive-ion electrospray ionization (ESI). Distinct peaks were observed for alendronate and for the internal standard on each channel within 1 min. The regression equations showed good linearity within the ranges of 2.0–100 ng/0.5 ml for the plasma and 1.0–100 ng/0.5 ml for the dialysate, with the limits of detection at 1.0 ng/0.5 ml for the plasma and 0.5 ng/0.5 ml for the dialysate. Extraction efficiencies for alendronate for the plasma and dialysate were 41.1–51.2% and 63.6–73.4%, respectively. The coefficient of variation (CV) was ≤8.5%. The method was successfully applied to the analyses of real plasma and dialysate samples derived after intravenous administration of alendronate.
更多
查看译文
关键词
Alendronate,Methylation,Plasma,Dialysate,A L-column2 ODS metal-free column,Metal-free HPLC-MS/MS