Accuracy of Electrode Placement in IRE Treatment with Navigated Guidance

CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY(2021)

引用 3|浏览0
暂无评分
摘要
Purpose Evaluate the accuracy of multiple electrode placements in IRE treatment of liver tumours using a stereotactic CT-based navigation system. Method Analysing data from all IRE treatments of liver tumours at one institution until 31 December 2018. Comparing planned with validated electrode placement. Analysing lateral and angular errors and parallelism between electrode pairs Results Eighty-four tumours were treated in 60 patients. Forty-six per cent were hepatocellular carcinoma, and 36% were colorectal liver metastases. The tumours were located in all segments of the liver. Data were complete from 51 treatments. Two hundred and six electrodes and 336 electrode pairs were analysed. The median lateral and angular error, comparing planned and validated electrode placement, was 3.6 mm (range 0.2–13.6 mm) and 3.1° (range 0°–16.1°). All electrodes with a lateral error >10 mm were either re-positioned or excluded before treatment. The median angle between the electrode pairs was 3.8° (range 0.3°–17.2°). There were no electrode placement-related complications. Conclusion The use of a stereotactic CT-based system for navigation of electrode placement in IRE treatment of liver tumours is safe, accurate and user friendly.
更多
查看译文
关键词
Liver,Ablation,IRE,Navigation,Stereotactic,Hepatocellular carcinoma
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要