A Transit Time-Resolved Microflow Cytometry-Based Agglutination Immunoassay For On-Site C-Reactive Protein Detection

MICROMACHINES(2021)

引用 3|浏览16
暂无评分
摘要
An accurate and rapid microflow cytometry-based agglutination immunoassay (MCIA) suitable for on-site antibody or antigen detection was proposed. In this study, quantitative C-reactive protein (CRP) detection was chosen as a model assay in order to demonstrate the detection principle. The average transit time was employed to estimate the extent of the agglutination reaction and improve the detection accuracy as compared to the intensity-dependent methods. The detection time was less than 8 min. and only a 20 mu L serum sample was needed for each test. The results showed a linear relationship between the average transit time of aggregates and CRP concentrations ranging from 0 to 1 mu g/mL. The R-2 of this relationship was 0.99. The detection limit of this technology was 0.12 mu g/mL CRP. The system used for CRP detection can be extended to also monitor other clinically relevant molecules.
更多
查看译文
关键词
C-reactive protein, microflow cytometry, agglutination immunoassay, on-site detection