Flat Panel Detector C-Arms Are Associated With Dramatically Reduced Radiation Exposure During Ureteroscopy And Produce Superior Images

JOURNAL OF ENDOUROLOGY(2021)

引用 1|浏览6
暂无评分
摘要
Background: We wished to determine whether newly available flat panel detector (FPD) c-arms were (1) associated with lower radiation dose during ureteroscopy (URS) than conventional image intensifier (CII) c-arms and (2) to compare fluoroscopic image quality between the units.Materials and Methods: We retrospectively reviewed 44 consecutive patients undergoing URS at a pediatric hospital, with c-arms assigned by availability in the operating room. We performed dosimetry experiments using the same c-arms on standard phantoms.Results: Patient and case characteristics did not differ significantly between the two groups of patients. The median dose in the FPD group was less than a quarter of the dose in the CII group, 0.48 [0.42, 0.97] mGy vs 2.2 [1.1, 3.8] mGy, p < 0.0001. The FPD dose remained at less than one-third of the CII dose accounting for any difference in fluoroscopy time, and remained significant in a multivariate model including fluoroscopy time and patient weight (beta = 2.4, p = 0.007). Phantom studies showed higher image quality for FPDs at all simulated patient sizes, even at lower radiation doses.Conclusions: This is the first report comparing radiation dose from c-arms of image intensifiers and FPDs in adults or children. Use of an FPD during URS was associated with a substantially decreased absorbed dose for patients while simultaneously improving image quality.
更多
查看译文
关键词
radiation, ureteroscopy, pediatrics, imaging, c-arm