Long-term and efficient operation of the MWPC muon detector at LHCb

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2020)

引用 0|浏览2
暂无评分
摘要
With its ∼1650m2 of MWPCs, the muon detector of LHCb is one of the largest instrument of this kind worldwide, and one of the most irradiated. Currently we run at the relatively low instantaneous luminosity of 4⋅1032 cm−2s−1, nevertheless the most irradiated MWPCs already integrated ∼0.7 C/cm of accumulated charge per wire. The statistics of gas gaps affected by high voltage trips in the proportional chambers is presented for the whole period of operation. Most of the problematic chambers were successfully recovered in situ during data taking, under the nominal LHC beam conditions, by means of a long-term HV training (with the working gas mixture). The appearing of self-sustained currents in one of the MWPC gaps and the effectiveness of the recovery procedures put in place, indicate that the large majority of the trips are due to Malter effect. The method has proven to be very effective, allowing to keep the muon detector efficiency very close 100%, as it was initially designed. In parallel, a test has been performed of a systematic addition of a small amount of oxygen to the nominal gas mixture: results of this test will be discussed.
更多
查看译文
关键词
Muon spectrometers,Wire chambers (MWPC),Malter effect