Dosimetric properties of a commercial PTW 60019 synthetic diamond detector in small photon fields

Xue Chang,Kun Wang, Jian Zhang,Zhipeng Wang,Sunjun Jin

Chinese Journal of Radiological Medicine and Protection(2018)

引用 0|浏览0
暂无评分
摘要
Objective To investigate the dosimetric properties of PTW 60019 synthetic diamond detector in small photon beams. Methods A PTW 60019 synthetic diamond detector was tested under 6 and 10 MV photon beams, respectively. Linearity with dose, dose rate dependence and off-axis ratio were measured and compared to those measured by an IBA SFD. Percentage depth doses were measured and compared to those measured by an IBA SFD and a PTW 31010 semiflex chamber. Total scatter factors were measured and compared to those measured by an IBA SFD and a PTW 31016 PinPoint chamber. Results The dose response of a PTW 60019 synthetic diamond detector showed a good linear behavior as a function of dose, with observed deviations below 0.2% over a dose range from 100 to 1 000 MU. The dose rate response was almost independent, with deviations below 0.2% in the dose rate range from 37 to 614 MU/min. For the fields of 20-100 mm in diameter, there were dose differences in percentage depth doses within 1% as compared to an IBA SFD and a PTW 31010 semiflex chamber. For the 10 mm diameter field, the differences were up to 5.8% in the build-up region. Off-axis ratios measurements showed a good agreement among the involved detectors (<1%). The higher differences appeared in the penumbra region. A good agreement was also found in terms of total scatter factor measurements for the related detectors. Conclusions The observed dosimetric properties of the PTW 60019 synthetic diamond detector indicate that it is a suitable candidate for small photon beam dosimetry.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要