Verification Of Cascade Optical Coherence Tomography For Freeform Optics Form Metrology

OPTICS EXPRESS(2021)

引用 1|浏览5
暂无评分
摘要
Freeform optical components enable dramatic advances for optical systems in both performance and packaging. Surface form metrology of manufactured freeform optics remains a challenge and an active area of research. Towards addressing this challenge, we previously reported on a novel architecture, cascade optical coherence tomography (C-OCT), which was validated for its ability of high-precision sag measurement at a given point. Here, we demonstrate freeform surface measurements, enabled by the development of a custom optical-relay-based scanning mechanism and a unique high-speed rotation mechanism. Experimental results on a flat mirror demonstrate an RMS flatness of 14 nm (similar to lambda/44 at the He-Ne wavelength). Measurement on a freeform mirror is achieved with an RMS residual of 69 nm (similar to lambda/9). The system-level investigations and validation provide the groundwork for advancing C-OCT as a viable freeform metrology technique. (C) 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement