Audit of Sweat Chloride Testing Reveals Analytical Errors.

Clinical chemistry and laboratory medicine(2021)

引用 0|浏览5
暂无评分
摘要
Abstract Objectives Sweat chloride testing (SCT) is the mainstay for the diagnosis of cystic fibrosis (CF) and biomarker in the evaluation of CFTR-modifying drugs. To be a reliable and valid tool, analytical variance (CVA) must be minimized. However, external quality assessments have revealed significant deviations in routine clinical practice. Our goal was to identify and quantify technical errors through proficiency testing and simulations. Methods Chloride concentrations of three blinded samples (each as triplicates) were measured in 9 CF centers using a chloridometer in a routine setting. Technical errors were simulated and quantified in a series of measurements. We compared imprecision and bias before and after a counseling session by evaluating coefficients of variation (CV), adherence to tolerance limits, and inter-rater variability coefficients. Results Pipetting errors resulting in changes in sample volume were identified as the main source of error with deviations up to 41%. After the counseling session, the overall CVA decreased from 7.6 to 5.2%, the pass rate increased from 67 to 92%, and the inter-rater variability diminished. Significant deviations continued to be observed in individual centers. Conclusions Prevention of technical errors in SCT decreases imprecision and bias. Quality assurance programs must be established in all CF centers, including staff training, standard operating procedures, and proficiency testing.
更多
查看译文
关键词
analytical error,chloridometer,cystic fibrosis,external quality assurance,laboratory proficiency testing,sweat chloride testing
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要