Comparison Of Simoa(Tm) And Ella(Tm) To Assess Serum Neurofilament-Light Chain In Multiple Sclerosis

ANNALS OF CLINICAL AND TRANSLATIONAL NEUROLOGY(2021)

引用 38|浏览6
暂无评分
摘要
We compared Simoa(TM) and Ella(TM) immunoassays to assess serum neurofilament-light chain levels in 203 multiple sclerosis patients from the OFSEP HD study. There was a strong correlation (rho = 0.86, p < 0.0001) between both platforms. The Ella(TM) instrument overestimated values by 17%, but as the data were linear (p = 0.57), it was possible to apply a correction factor to Ella(TM) results. As for Simoa(TM), serum neurofilament-light chain levels measured by Ella(TM) were correlated with age and EDSS and were significantly higher in active multiple sclerosis, suggesting that these assays are equivalent and can be used in routine clinical practice.