Low Temperature Thomson Scattering On Mast-U

REVIEW OF SCIENTIFIC INSTRUMENTS(2021)

引用 7|浏览5
暂无评分
摘要
A new divertor Thomson scattering system has been developed for the MAST-U tokamak. The diagnostic will produce electron density and temperature profiles along the Super-X strike leg. The existing polychromator design has been adapted for low temperature measurements. A new 1061 nm channel with 2 nm bandwidth has been added to enable measurements down below the previous similar to 5 eV limit on the core system. The optical filters used in the system have OD6 light rejection alongside a 1064.1 nm laser line filter to reduce stray light in the digitized channels. A new averaging technique has been applied to the scattered signal traces to improve the core Thomson data in the scrape-off layer. The technique reduces the systematic noise level in this region. This leads to a reduction in the error values for electron density and temperature measurements and, in particular, the digitizer noise. The technique has been applied to produce a radial profile for a number of L-mode MAST discharges down to very low densities of similar to 1 x 10(18) m(-3).
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper