Sharing Mechanical Ventilator: In Vitro Evaluation Of Circuit Cross-Flows And Patient Interactions

MEMBRANES(2021)

引用 3|浏览16
暂无评分
摘要
During the COVID-19 pandemic, a shortage of mechanical ventilators was reported and ventilator sharing between patients was proposed as an ultimate solution. Two lung simulators were ventilated by one anesthesia machine connected through two respiratory circuits and T-pieces. Five different combinations of compliances (30-50 mL x cmH(2)O(-1)) and resistances (5-20 cmH(2)O x L-1 x s(-1)) were tested. The ventilation setting was: pressure-controlled ventilation, positive end-expiratory pressure 15 cmH(2)O, inspiratory pressure 10 cmH(2)O, respiratory rate 20 bpm. Pressures and flows from all the circuit sections have been recorded and analyzed. Simulated patients with equal compliance and resistance received similar ventilation. Compliance reduction from 50 to 30 mL x cmH(2)O(-1) decreased the tidal volume (V-T) by 32% (418 +/- 49 vs. 285 +/- 17 mL). The resistance increase from 5 to 20 cmH(2)O x L-1 x s(-1) decreased V-T by 22% (425 +/- 69 vs. 331 +/- 51 mL). The maximal alveolar pressure was lower at higher compliance and resistance values and decreased linearly with the time constant (r(2) = 0.80, p < 0.001). The minimum alveolar pressure ranged from 15.5 +/- 0.04 to 16.57 +/- 0.04 cmH(2)O. Cross-flows between the simulated patients have been recorded in all the tested combinations, during both the inspiratory and expiratory phases. The simultaneous ventilation of two patients with one ventilator may be unable to match individual patient's needs and has a high risk of cross-interference.
更多
查看译文
关键词
COVID-19, SARS-CoV2 infection, cross-flow dynamics, shared ventilation, sharing mechanical ventilator, two patients one ventilator
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要