Reliability Of Ultrasound Hepatorenal Index And Magnetic Resonance Imaging Proton Density Fat Fraction Techniques In The Diagnosis Of Hepatic Steatosis, With Magnetic Resonance Spectroscopy As The Reference Standard

PLOS ONE(2021)

引用 3|浏览5
暂无评分
摘要
Purpose To evaluate the reliability of ultrasound hepatorenal index (US-HRI) and magnetic resonance imaging proton density fat fraction (MRI-PDFF) techniques in the diagnosis of hepatic steatosis, with magnetic resonance spectroscopy proton density fat fraction (MRS-PDFF) as the reference standard. Materials and methods Fifty-two adult volunteers (30 men, 22 women; age, 31.5 +/- 6.5 years) who had no history of kidney disease or viral/alcoholic hepatitis were recruited to undergo abdominal US, MRI, and MRS examinations. US-HRI was calculated from the average of three pairs of regions of interest (ROIs) measurements placed in the liver parenchyma and right renal cortex. On MRI, the six-point Dixon technique was employed for calculating proton density fat fraction (MRI-PDFF). An MRS sequence with a typical voxel size of 27 ml was chosen to estimate MRS-PDFF as the gold standard. The data were evaluated using Pearson's correlation coefficient and receiver operating characteristic (ROC) curves. Results The Pearson correlation coefficients of US-HRI and MRI-PDFF with MRS-PDFF were 0.38 (p = 0.005) and 0.95 (p<0.001), respectively. If MRS-PDFF >= 5.56% was defined as the gold standard of fatty liver disease, the areas under the curve (AUCs), cut-off values, sensitivities and specificities of US-HRI and MRI-PDFF were 0.74, 1.54, 50%, 91.7% and 0.99, 2.75%, 100%, 88.9%, respectively. The intraclass correlation coefficients (ICCs) of US-HRI and MRI-PDFF were 0.70 and 0.85. Conclusion MRI-PDFF was more reliable than US-HRI in diagnosing hepatic steatosis.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要