Continuous Glucose Monitoring in the Management of Neonates With Persistent Hypoglycemia and Congenital Hyperinsulinism

JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM(2022)

引用 19|浏览6
暂无评分
摘要
Background Persistent hypoglycemia is common in the newborn and is associated with poor neurodevelopmental outcome. Adequate monitoring is critical in prevention, but is dependent on frequent, often hourly blood sampling. Continuous glucose monitoring (CGM) is increasingly being used in children with type 1 diabetes mellitus, but use in neonatology remains limited. We aimed to introduce real-time CGM to provide insights into patterns of dysglycemia and to support the management of persistent neonatal hypoglycemia. Methods This is a single-center retrospective study of real-time CGM use over a 4-year period in babies with persistent hypoglycemia. Results CGMs were inserted in 14 babies: 8 term and 6 preterm infants, 9 with evidence of congenital hyperinsulinism (CHI). A total of 224 days of data was collected demonstrating marked fluctuations in glucose levels in babies with CHI, with a higher sensor glucose SD (1.52 +/- 0.79 mmol/L vs 0.77 +/- 0.22 mmol/L) in infants with CHI compared with preterm infants. A total of 1254 paired glucose values (CGM and blood) were compared and gave a mean absolute relative difference of 11%. Conclusion CGM highlighted the challenges of preventing hypoglycemia in these babies when using intermittent blood glucose levels alone, and the potential application of CGM as an adjunct to clinical care.
更多
查看译文
关键词
Glucose, hyperinsulinism, hypoglycemia, continuous glucose monitoring
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要