Minimal Artifact Actively Shimmed Metallic Needles in MRI

Magnetic resonance in medicine(2021)

引用 2|浏览15
暂无评分
摘要
Purpose: Signal voids caused by metallic needles pose visualization and monitoring challenges in many MRI applications. In this work, we explore a solution to this problem in the form of an active shim insert that fits inside a needle and corrects the field disturbance (Delta B-0) caused by the needle outside of it. Methods: The Delta B-0 induced by a 4 mm outside-diameter titanium needle at 3T is modeled and a two-coil orthogonal shim set is designed and fabricated to shim the Delta B-0. Signal recovery around the needle is assessed in multiple orientations in a water phantom with four different pulse sequences. Phase stability around the needle is assessed in an ex-vivo porcine tissue dynamic gradient echo experiment with and without shimming. Additionally, heating of the shim insert is assessed under 8 min of continuous operation with 1A current and concurrent imaging. Results: An average recovery of similar to 63% of lost signal around the needle across orientations is shown with active shimming with a maximum current of 1.172 A. Signal recovery and correction of the underlying Delta B-0 is shown to be independent of imaging sequence. Needle-induced phase gradients outside the perceptible signal void are also minimized with active shimming. Temperature rise of up to 0.9 degrees Celsius is noted over 8 min of continuous IA active shimming operation. Conclusion: A sequence independent method for minimization of metallic needle induced signal loss using an active shim insert is presented. The method has potential benefits in a range of qualitative and quantitative interventional MRI applications.
更多
查看译文
关键词
active shimming,field inhomogeneity,interventional MRI,metallic needles,shim coils
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要