The lack of organs available for transplantation is a global problem. The high mortality rates on the waiting list and the high number of discarded livers are reasons to develop new tools in the preservation and transplantation process. New tools should also be available for low-income countries. This article reports the development of customized normothermic machine perfusion (NMP). An ex vivo dual perfusion machine was designed, composed of a common reservoir organ box (CRO), a centrifugal pump (portal system, low pressure), and a roller pump (arterial system, high pressure). Porcine livers (n = 5) were perfused with an oxygenated normothermic (37.) strategy for 3 hours. Hemodynamic variables, metabolic parameters, and bile production during preservation were analyzed. Arterial and portal flow remain stable during perfusion. Total bilirubin production was 11.25 mL (4-14.5) at 180 minutes. The median pH value reached 7.32 (7.25-7.4) at 180 minutes. Lactate values decreased progressively to normalization at 120 minutes. This perfusion setup was stable and able to maintain the metabolic activity of a liver graft in a porcine animal model. Design and initial results from this customized NMP are promising for a future clinical application in low-income countries.