Improved Imaging Using Mn He-Alpha X Rays At Omega Ep

REVIEW OF SCIENTIFIC INSTRUMENTS(2021)

引用 1|浏览68
暂无评分
摘要
In this paper, we report on a crystal based x-ray imaging system fielded at the OMEGA EP laser facility. This new system has a pointing accuracy of +/100 mu m, a temporal resolution down to 100 ps (depending on backlighter characteristics), variable magnification, and a spatial resolution of 21.9 mu m at the object plane at a magnification of 15x. The system is designed to use a crystal along the crystal plane that satisfies the Bragg condition for the x ray of interest. The thin crystal is then bent into a spherical geometry and attached to a glass backing substrate to hold it in the diagnostic, and the x rays are imaged onto a charge coupled device. We report on data acquired with the new Los Alamos National Laboratory supplied spherical quartz crystal to image the Mn He-alpha 6.15 keV line emission.& nbsp;Published under an exclusive license by AIP Publishing.