Contrast Agents For X-Ray Luminescence Computed Tomography

APPLIED OPTICS(2021)

引用 5|浏览15
暂无评分
摘要
Imaging probes are an important consideration for any type of contrast agent-based imaging method. X-ray luminescence imaging (XLI) and x-ray luminescence computed tomography (XLCT) are both contrast agent-based imaging methods that employ x-ray excitable scintillating imaging probes that emit light to be measured for optical imaging. In this work, we compared the performance of several select imaging probes, both commercial and self-synthesized, for application in XLI/XLCT imaging. Commercially available cadmium telluride quantum dots (CdTe QDs) and europium-doped gadolinium oxysulfide (GOS:Eu) microphosphor as well as synthesized NaGdF4 nanophosphors doped with either europium or terbium were compared through their x-ray luminescence emission spectra, luminescence intensity, and also by performing XLCT scans using phantoms embedded with each of the imaging probes. Each imaging probe displayed a unique emission spectrum that was ideal for deep-tissue optical imaging. In terms of luminescence intensity, due to the large particle size, GOS:Eu had the brightest emission, followed by NaGdF4 : Tb, NaGdF4 : Eu, and finally the CdTe QDs. Lastly, XLCT scans showed that each imaging probe could be reconstructed with good shape and location accuracy. (C) 2021 Optical Society of America