Intraoperative Invasive Blood Pressure Monitoring And The Potential Pitfalls Of Invasively Measured Systolic Blood Pressure

Sean Lam, Hong Liu, Zhongping Jian,Jos Settels,Christian Bohringer

CUREUS(2021)

引用 7|浏览0
暂无评分
摘要
Invasive intraarterial blood pressure measurement is currently the gold standard for intraoperative hemodynamic monitoring but accurate systolic blood pressure (SBP) measurement is difficult in everyday clinical practice, mostly because of problems with hyper-resonance or damping within the measurement system, which can lead to erroneous treatment decisions if these phenomena are not recognized. A hyperresonant blood pressure trace significantly overestimates true systolic blood pressure while underestimating the diastolic pressure. Invasively measured systolic blood pressure is also significantly more affected than mean blood pressure by the site of measurement within the arterial system. Patients in the intraoperative period should be treated based on the invasively measured mean blood pressure rather than the systolic blood pressure. In this review, we discuss the pros/cons, mechanisms of invasive blood pressure measurements, and the interpretation of the invasively measured systolic blood pressure value.
更多
查看译文
关键词
invasive intraarterial blood pressure, hyperresonance, damping, hemodynamics, monitoring