谷歌浏览器插件
订阅小程序
在清言上使用

Physical Confirmation of Biology-Guided Radiotherapy Directed at Static Targets with Varying Shapes and Background Contrast Environments.

International journal of radiation oncology, biology, physics(2021)

引用 0|浏览8
暂无评分
摘要
Dose accuracy results indicate that BgRT plans can be delivered accurately over a variety of test conditions (different target shapes, attenuating media and activity concentration levels) to FDG-avid targets, with performance similar to SBRT treatments.