SHEAR-WAVE DISPERSION SLOPE OF THE LIVER: EFFECT OF STUDY PROTOCOL AND ASCITES ON THE MEASUREMENT APPLICABILITY

Xiao-Qing Zhang,Jie Zeng,Jie-Yang Jin,Jin-Fen Wang, Yu-Ying Chi,Rong-Qin Zheng

ULTRASOUND IN MEDICINE AND BIOLOGY(2022)

引用 10|浏览4
暂无评分
摘要
This study aimed to evaluate the shear-wave dispersion (SWD) scanning protocol including the minimum number of measurements and better size of the region of interest (ROI), as well as the influence of ascites on the measurement applicability. Patients who had undergone serial SWD examinations between July 2019 and December 2020 were included. In patients with chronic liver disease (group A), two different ROI sizes were applied, and at least 10 measurements were repeated to determine the minimum number of measurements and better ROI size. In patients with liver failure (group B), failure and unreliable results were compared between patients with and without ascites. A minimum of five measurements when using a 20-mm ROI and six measurements when using a 10-mm ROI were required. Compared with using a 20-mm ROI, a 10-mm ROI showed a higher unreliable rate. The failure and unreliable rates of SWD in patients with ascites were significantly higher than those in patients without ascites. SWD examination required at least five measurements when using a 20-mm ROI and six measurements when using a 10-mm ROI. A larger ROI was associated with higher reliability, and ascites influenced the failure and reliability of the SWD measurement. (C) 2021 World Federation for Ultrasound in Medicine & Biology. All rights reserved.
更多
查看译文
关键词
Dispersion, Shear-wave elastography, Liver, Scanning protocol, Applicability