Clinical performance of implantable cardioverter-defibrillator lead monitoring diagnostics

Heart Rhythm(2022)

引用 4|浏览34
暂无评分
摘要
BACKGROUND Implantable cardioverter-defibrillator (ICD) lead monitoring diagnostic alerts facilitate the diagnosis of structural lead failure.OBJECTIVE The purpose of this study was to prospectively study the performance of Medtronic ICD lead monitoring alerts.& nbsp;METHODS A prespecified ancillary substudy, World-Wide Randomized Antibiotic Envelope Infection Prevention Trial, was conducted in patients with an ICD with all available alerts enabled. The investigators reported possible lead system events (LSEs), with or without an alert. An independent committee reviewed all data and classified events as lead failure, other LSE, or nonlead system events (NLEs).& nbsp;RESULTS In 4942 patients who were followed for 19.4 6 8.7 months, there were 124 alerts (65 LSEs, 59 NLEs) and 19 LSEs without an alert. Lead monitoring alerts had 100% sensitivity for the 48 adjudicated lead failures (95% confidence interval 92.6%-100%) and for 10 events adjudicated as either lead failure or connection issue. The positive predictive value of alerts for lead failure was 38.7% (48 of 124). For 34 pace-sense lead failures, an alert that incorporated oversensing was more sensitive than the pacing impedance threshold alert (33 patients [97.1%] vs 9 patients [26.5%]; P , .0001). However, the sensitivity was only 13.6% for lead dislodgments or perforations. Inappropriate shocks occurred in 2 patients with pace-sense lead failure (5.9%). No patient had unnecessary lead replacement for any of the NLEs.& nbsp;CONCLUSION In this first real-world prospective study, lead monitoring alerts had 100% sensitivity for identifying lead failures. Although their positive predictive value was modest, no false-positive alerts resulted in an unnecessary lead replacement. For the diagnosis of pace-sense lead failure, an alert for over sensing was more sensitive than a pacing impedance threshold alert. (C) 2021 Heart Rhythm Society.& nbsp;
更多
查看译文
关键词
Defibrillator,Lead failure,Lead monitoring,Remote monitoring,Ventricular tachycardi
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要