Cell-Tak coating may cause mis-normalization of Seahorse metabolic flux data

biorxiv(2021)

引用 0|浏览2
暂无评分
摘要
Metabolic flux investigations of cells and tissue samples are a rapidly advancing tool in diverse research areas. Reliable methods of data normalization are crucial for an adequate interpretation of results and to avoid a misinterpretation of experiments and incorrect conclusions. The most common methods for metabolic flux data normalization are to cell number, DNA and protein. Data normalization may be affected by a variety of factors, such as density, healthy state, adherence efficiency, or proportional seeding of cells. The mussel-derived adhesive Cell-Tak is often used to immobilize poorly adherent cells. Here we demonstrate that this coating may strongly affect the fluorescent detection of DNA leading to an incorrect and highly variable normalization of metabolic flux data. Protein assays are much less affected and cell counting can virtually completely remove the effect of the coating. Cell-Tak coating also affects cell shape in a cell line-specific manner and may change cellular metabolism. Based on these observations we recommend cell counting as a gold standard normalization method for Seahorse metabolic flux measurements with protein content as a reasonable alternative. ### Competing Interest Statement The authors have declared no competing interest.
更多
查看译文
关键词
metabolic,cell-tak,mis-normalization
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要