Nanopore ion sources deliver single amino acid and peptide ions directly into high vacuum

bioRxiv (Cold Spring Harbor Laboratory)(2022)

引用 0|浏览1
暂无评分
摘要
We report a nanopore ion source for mass spectrometry that delivers ions directly into high vacuum from aqueous solutions. The ion source comprises a pulled quartz capillary with a sub-100 nm opening. Ions escape an electrified meniscus between the liquid and high vacuum by ion evaporation and travel along collisionless trajectories to the ion detector. We measure mass spectra of 16 different amino acid ions as well as post-translationally modified variants of glutathione in unsolvated states using the nanopore ion source. The emitted current is composed of ions rather than charged droplets, and more than 90% of the current can be recovered in a distant collector. The nanopore ion source could enable more sensitive proteomic analyses. It circumvents the sample loss mechanisms inherent to conventional electrospray ionization (ESI), where charged droplets are sprayed into a background gas that scatters ions and degrades their transmission. ### Competing Interest Statement DS declares a competing interest in Oxford Nanopore Technologies, Ltd.
更多
查看译文
关键词
nanopore ion source,peptide ions,single amino acid,gas phase