Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years

SCIENTIFIC REPORTS(2017)

引用 23|浏览10
暂无评分
摘要
This study evaluated the performance of plusoptiX A12C in detecting amblyopia risk factors (ARFs) in Chinese children aged 3-to-4-year. PlusoptiX examination was successfully conducted among 1,766 subjects without cycloplegia to detect refractive error, asymmetry and media opacity. Cycloplegic retinoscopy (CR) was conducted on 357 children suspected of having vision abnormalities. Statistical differences between CR and the device were confirmed using the mean spherical value (+1.41 ± 0.87 D versus +1.14 ± 0.81 D), cylindrical value (−0.47 ± 0.64 versus −0.84 ± 0.78) and spherical equivalent (SE) value (+1.17 ± 0.84 D versus +0.72 ± 0.64 D) (all P < 0.0001). In the emmetropia group, the differences were statistically significant for the cylinder and SE (all P < 0.0001) but not the sphere ( P = 0.33). In the hyperopia group, the differences were statistically significant for the sphere, cylinder and SE (all P < 0.0001). For refractive and strabismic ARFs detection, the sensitivity, specificity, positive predictive value, and negative predictive value were calculated, respectively.
更多
查看译文
关键词
Paediatric research,Refractive errors,Science,Humanities and Social Sciences,multidisciplinary