Extended Utilization Of A Ge-68/Ga-68 Reference Source As A Mock F-18 Source In The Ionization Chamber Calibration

APPLIED RADIATION AND ISOTOPES(2018)

引用 1|浏览2
暂无评分
摘要
Possible extended utilization of a Ge-68/Ga-68 source as a surrogate F-18 reference source was studied. Some effects and uncertainties might arise from the difference of constructions between chambers that were used in the source calibration and those practically in use. To extend utilization of the surrogate source, response ratio of real F-18 to Ge-68/Ga-68 sources were determined for three models of chambers. Monte-Carlo simulation using EGS5 code was also carried out to evaluate uncertainties expected. The present result shows that such a surrogate source would adapt to calibration even for chambers have with different construction from the reference chamber used as the determination of F-18 equivalent activity of the Ge-68/Ga-68 surrogate source.
更多
查看译文
关键词
Radiopharmaceuticals, Dose calibrator, Positron Emission Tomography (PET), Radioactivity measurement