Primary Standardization Of Ga-67 Radiopharmaceuticals

APPLIED RADIATION AND ISOTOPES(2008)

引用 3|浏览0
暂无评分
摘要
The two-dimensional extrapolation technique and dead time extrapolation technique of 4 pi beta-gamma-coincidence counting were used to standardize the activity of Ga-67. The counting results of the two absolute counting techniques showed good agreement while the two-dimensional extrapolation technique could save about 50% of the total counting time in this study. The accuracy of the Capintec CRC-15R radionuclide calibrator used by the radiopharmacy of INER was studied in this research. A new calibration setting number, 99, was recommended to the radiopharmacy of INER in the Bayer 10 mL sterile empty vial geometry. (c) 2008 Elsevier Ltd. All rights reserved.
更多
查看译文
关键词
Ga-67,standardization,coincidence counting,Capintec CRC-15R calibration setting