Interference By Piperacillin/Tazobactam In The Measurement Of Creatinine With The Jaffe Method And Of Total Protein With The Biuret Method

CRITICAL CARE AND RESUSCITATION(2021)

引用 0|浏览5
暂无评分
摘要
Serum creatinine and total protein are routinely measured biochemical parameters used in clinical medicine. An abnormal result caused by interference with the assay does not accurately reflect a patient's clinical state and therefore risks misleading clinicians.We report the case of a patient who had unexplainable high creatinine and total protein results. The blood collection was contaminated with intravenous fluid and the patient was receiving piperacillin/tazobactam. Additional laboratory studies demonstrated piperacillin/tazobactam was the cause of the false positive results and the elevation in both serum creatinine and protein level was dependent on the concentration of antibiotic present.
更多
查看译文
关键词
creatinine,piperacillin/tazobactam,total protein,jaffe method