Magnetic Resonance Imaging compatible Elastic Loading Mechanism (MELM): A minimal footprint device for MR imaging under load

2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC)(2021)

引用 1|浏览3
暂无评分
摘要
Quantitative Magnetic Resonance Imaging (MRI) can enable early diagnosis of knee cartilage damage if imaging is performed during the application of load. Mechanical loading via ropes, pulleys and suspended weights can be obstructive and require adaptations to the patient table. In this paper, a new lightweight MRI-compatible elastic loading mechanism is introduced. The new device showed sufficient linearity (vertical bar alpha/beta vertical bar= 0.42 +/- 0.25), reproducibility (CoV = 5 +/- 2%), and stability (CoV = 0.5 +/- 0.1%). In vivo and ex vivo scans confirmed the ability of the device to exert sufficient force to study the knee cartilage under loading conditions, inducing up to a 29% decrease in T-2* of the central medial cartilage. With this device mechanical loading can become more accessible for researchers and clinicians, thus facilitating the translational use of MRI hiomarkers for the detection of cartilage deterioration.
更多
查看译文
关键词
Humans,Magnetic Resonance Imaging,Reproducibility of Results