Two pumps or one pump? A comparison of human liver normothermic machine perfusion devices for transplantation

ARTIFICIAL ORGANS(2022)

引用 4|浏览17
暂无评分
摘要
Background Normothermic machine perfusion provides continuous perfusion to ex situ hepatic grafts through the portal vein and the hepatic artery. Because the portal vein has high flow with low pressure and the hepatic artery has low flow with high pressure, different types of perfusion machines have been employed to match the two vessels' infusion hemodynamics. Methods We compared transplanted human livers perfused through a 2-pump (n = 9) versus a 1-pump perfusion system (n = 6) where a C-clamp is used as a tubing constrictor to regulate hemodynamics. Results There was no significant difference between groups in portal vein or hepatic artery flow rate. The 1-pump group had more hemoglobin in the perfusate. However, there was no significant difference in plasma hemoglobin between the 2-pump and 1-pump groups at each time point or in the change in levels, proving no hemolysis occurred due to C-clamp tube constriction. After transplantation, the 2-pump group had two cases of early allograft dysfunction (EAD), whereas the 1-pump group had no EAD. There was no graft failure or patient death in either group during follow-up ranging from 20-52 months. Conclusions Our data show that the 1-pump design provided the same hemodynamic output as the 2-pump design, with no additional hemolytic risk, but with the benefits of lower costs, easier transport and faster and simpler setting.
更多
查看译文
关键词
hemolysis, liver, normothermic machine perfusion, transplantation